InTemp® CX600 Dry Ice/CX700 Cryogenic Logger Manual
22025-G
InTemp CX600 Dry Ice and CX700 Cryogenic loggers are designed for cold shipment monitoring
and have a built-in external probe that can measure temperatures as low as -95°C (-139°F) for
the CX600 series or -200°C (-328°F) for the CX700
series. The loggers include a protective
sheath to prev
ent cuttin
g the cable during shipme
nt and a clip for mountin
g the probe.
Designed for wireless communica
ti
on with a mobile device, these Bluetooth® Low Energy-
enabled loggers use the InTemp app, and InTempConnect® web-based
softwa
re to form the
InTemp temperature monitoring solution. Using
the InTemp app on your phone or tablet, you
can configure the loggers
and then download them to share and view logger reports, which
include logged data, excursions, and alarm info
rmation. Or, you
can use InTempConnect to
configure and
download CX series loggers via the CX5000 Gateway. The InTempVerify™ app is
also available to easily download loggers and
automatically uploa
d reports to InTempConnect.
Onc
e logged data is uploaded to InTempConnec
t, you can view logger configurations, build
custom reports, monitor trip information, a
nd more. Both the CX600 and CX700 series loggers
are available in single-use 90-day models (CX602 and CX702) or m
ultiple-use 365-day models
(CX603 or CX703).
Specifications
Temperature Probe
CX602 & CX603
CX702 & CX703
Sensor Range
-95° to 50°C (-139° to 122°F)
-200° to 50°C (-328° to 122°F);
To ensure insula
ti
on integrity,
avoid
fl
exing the wire during
exposure to temperatures below
-100°C (-148°F)
Accuracy
±1.5°C from -95° to -40°C (±2.7°F
from -139° to -40°F)
±1.0°C from -40° to -10°C (±1.8°F
from -40° to 14°F)
±0.5°C from -10° to 50°C (±0.9°F
from -14° to 122°F)
±1.5°C from -200° to -90°C (±2.7°F
from -328° to -130°F)
±1.0°C from -90° to -10°C (±1.8°F
from -328° to 14°F)
±0.5°C from -10° to 50°C (±0.9°F
from 14° to 122°F)
Resolution
0.013°C at 25°C (0.023°F at 77°F)
Drift
<0.1°C (0.18°F) per year
NIST Calibration
Three-point NIST calibra
ti
on,
probe only
CX702 and CX703: Single-point
NIST calibr
ation
, probe only
CX703-
UN
: Calibra
ti
on not
included
Cable Length
1.5 m (4.92
ft
)
Probe Dimensions
Food-grade stainless steel probe with
blunt tip
, 31.75 mm (1.25 inches)
long, 4.78 mm (0.188 inches) diameter
Logger
Logger Operating Range
-30° to 50°C (-22° to 122°F)
Radio Power
1 mW (0 dBm)
Transmission Range
Approximately 30.5 m (100
ft
) line-of-sight
Wireless Data Standard
Bluetooth Low Energy (Bluetooth Smart)
Type
CX602 and CX702: Single use
CX603, CX703, and CX703-
UN
: Mu
lti
ple use
Duration
CX602 and CX702: 90 days
CX603, CX703, and CX703-
U
N:
Up
to 1 year
Shelf Life
U
p to 1 year
Logging Rate
1 second to 18 hours
Time Accuracy
±1 minute per month at 25°C (77°F)
Battery
CR2450 3V lithium, not replaceable. Ba
tt
ery for CX602 and CX702
models lasts for dura
tion
speci
fi
ed when logger is used within the 1-year
shelf life. Ba
tt
ery life for the CX603 and CX703 models is 1 year, typical
with logging interval of 1 minute. Faster logging intervals, remaining
connected with the InTemp app, excessive downloads, and paging may
impact
batt
ery life.
(CX602 model shown)
InTemp CX600/CX700 and
Series Loggers
Models:
•
CX602, 90-day logger,
single use
•
CX603, 365-day logger,
mul
tipl
e use
•
CX702, 90-day logger,
single use
•
CX703, 365-day logger,
mul
tipl
e use
•
CX703-
U
N, 365-day logger,
mul
tipl
e use, without NIST
calibra
tio
n
Required Items:
•
InTemp app
•
Device with iOS
or
Android™ and Bluetooth