InTemp CX600 Dry Ice/CX700 Cryogenic Logger Manual 

 

www.prosensor.com

 

4

3.

 

Tap Set Logger Passkey. 

4.

 

Type a passkey up to 10 characters. 

5.

 

Tap Save. 

6.

 

Tap Disconnect. 

Only the phone or tablet used to set the passkey can then 

connect to the logger without entering a passkey; all other 

mobile devices will be required to enter the passkey. For 

example, if you set the passkey for the logger with your tablet 

and then try to connect to the device later with your phone, 

you will be required to enter the passkey on the phone but not 

with your tablet. Similarly, if othe

rs attempt to connect to the 

logger with different devices, th

en they would also be required 

to enter the passkey. To reset a passkey, connect to the logger, 

tap Set Logger Passkey, and select Reset to Factory Default. 

Downloading the Logger 

You can download the logger to a phone or tablet and generate 

reports that include logged data 

and alarm information. 

Reports can be shared immediately upon download or accessed 

later in the InTemp app. 

InTempConnect users:

 Privileges are required to download, 

preview, and share reports in the InTemp app. Report data is 

automatically u

ploaded to InTempConnect when you download 

the logger. Log in to InTempConnect to build custom reports 

(requires privileges). In a

ddition

, InTempConnect users can also 

download CX logge

rs automatically

 on a regular basis using the 

CX5000 Gateway. Or, if the logger 

was configured with a logger 

profil

e in which InTempVerify was enabled, then anyone can 

download the logger with the InTempVerify app. For details on 

the gateway and InTempVerify, see www.intempconnect/help. 
To download the logger with the InTemp app: 
1.

 

Press the button on the logger t

o wake it up. 

2.

 

Tap the Devices icon and connect to the logger. 

3.

 

Tap Download. 

4.

 

Choose a downlo

ad option: 

Important: 

CX602 and CX702 loggers cannot be 

restarted. If you want a CX602 or CX702 logger to 

con

ti

nue logging

 aft

er the download is complete, select 

Download & Continue. 

 

Download & Co

ntinu

e. The logger will 

continu

e logging 

once the download is complete. 

 

Download & Restart (CX603 models only). The logger will 

start a new data set using 

the same profile once the 

download is complete. Note that if the logger was 

originally 

configured with a pu

s

h button start, you must 

push the 

start button for logging to restart. 

 

Download & Stop. The logger will stop logging once the 

download is complete. 

A report of the download is generated and is also uploaded to 

InTempConnect if you are logged in to the InTemp app with 

your InTempConnect user creden

ti

als. 

In the app, tap Settings to ch

ange the default report type 

(Secure PDF or XLSX) and report 

sharing options.

 The report is 

also available in in both formats for sharing at a later time. Tap 

the Reports icon to access previously downloaded reports. See 

www.intempconnect.com/help for details on working with 

reports in both the InTemp app and InTempConnect. 

Logger Events 

The logger records the following events to track logger 

ope

ration and s

tatus. These events are listed in reports 

downloaded from the logger. 

Event Name 

Definition 

Configur

ed 

The logger was co

nfigur

ed by a user. 

Connected 

The logger was connected to the InTemp 

app. 

Started 

The logger started logging. 

Stopped 

The logger stopped logging. 

Downloaded 

The logger was downloaded. 

Alarm Out of 

Range/In Range 

An alarm has occurred because the reading 

was outside the alarm limits or back within 

range. 

Note:

 Although the reading may 

return to a normal range, the alarm 

indicator will not clear in the InTemp app 

and the alarm LED will con

ti

nue to blink.  

Safe Shutdown 

The battery 

level dropped below a safe 

oper

atin

g voltage and performed a safe 

shutdown. 

Deploying the Logger 

Use the mounting loop on the logger to secure it to a shipment 

or other application you are mo

nitoring. You can also remove 

the backing on the tape that is adhered to the to

p and bottom 

of the logger to 

mount it on a flat surface

Place the stainless steel probe in 

the plastic clip i

ncluded with 

the logger and clip it to a box or other item. 
The external probe cable has a protec

ti

ve sheath. Move the 

sheath as needed to posi

tion

 it where the cable will be 

protected during shipments from inadvertent cuts. 

Protecting the Logger 

Note:

 

Static electricity may cause the logger to stop logging.

 

The logger has been tested to 8 KV, but avoid electr

ostati

discharge by grounding yourself to protect the logger. For more 

information, s

e

arch for “static d

ischarge” on onsetcomp.com. 

Battery Information 

The logger uses 

one CR2450 non-replaceable lithium battery. 

Batt

ery life is not guaranteed past the 1-year logger shelf life. 

Battery life for the CX603 and

 CX703 models is 1 year, typical 

with a logging interval of 1 minute. Expected ba

tt

ery life for the 

CX603 and CX703 models vary based on the ambient 

temperature where the logger is deployed and the frequency of 

connections, downloads, and p

aging. Deployments in extremely 

cold or hot temperatures or a logging interval faster than 1 

minute can impact battery life. Estimates are n

ot guaranteed 

due to uncer

tainti

es

 in initi

al ba

tt

er

y conditions and op

er

atin

environment. 

 

 