RECOMMENDED SEPARATION DISTANCES BETWEEN 

PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE DEVICE

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. 

The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum 

distance between portable and mobile RF communications equipment (transmitters) and as recommended below, 

according to the maximum output power of the communications equipment.

Output Power of 

Transmitter in Watt

Separation Distance According To Frequency Of Transmitter In Meter

150 kHz to 80 MHz 

d = 1.2 √P

80 MHz to 800 MHz 

d = 1.2 √P

800 MHz to 2.5 GHz 

d = 2.3 √P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

TABLE 6: 

Summary of Contents for Dual Channel TENS

Page 1: ...User Manual Dual Channel TENS Prescription Strength Pain Relief Quick and Easy Set Up 20 Minute Treatment Timer 12 Preset Modes 100 Intensity Levels Treatment Record Log Auto Shut Off...

Page 2: ...ts may however be published in new editions of this manual All Rights Reserved Rev V2 0 2017 Please Note Compass Health Brands recommends consulting a physician before using a TENS unit to verify any...

Page 3: ...Programs 20 Intensity Adjustment 22 Start Treatment 22 Pause or Stop Treatment 23 Other Important Features 24 Safety Lock 24 Treatment Record Log 25 Electrode Positioning 26 Cleaning and Storage 29 Cl...

Page 4: ...tion Please read the complete manual carefully before using the device for the first time and keep this instruction manual in a convenient place or store with the device for future reference What is T...

Page 5: ...n Refer to diagrams on pages 26 28 for the ideal locations to place the electrodes to treat the most common forms of pain For other areas of pain place the electrodes on either side of the pain area P...

Page 6: ...must not be used in combination with the following medical devices Internally transplanted electronic medical devices such as pacemakers Electronic life support equipment such as respirators Electron...

Page 7: ...erate properly when the electrical stimulation device is in use If there are any underlying skin conditions i e erythema psoriasis eczema etc On open wounds or rashes over swollen red infected inflame...

Page 8: ...rning telling us that something is wrong Therefore if you suffer from any serious illness consult your physician in order to confirm that it is advisable for you to use this TENS unit Consult a physic...

Page 9: ...ainer when not in use Do not apply ointment or any solvent to the electrodes or to your skin because it will keep the electrode from functioning properly The electrode is already pre gelled and will a...

Page 10: ...TENS unit is not functioning properly or you feel discomfort immediately stop using the device Do not use for any other purpose except as described in this manual Do not insert the electrode plug int...

Page 11: ...inal source or cause of the pain including headache TENS is not a substitute for pain medications and other pain management therapies TENS devices do not cure diseases or injuries TENS is a symptomati...

Page 12: ...hildren The unit contains small pieces that may be swallowed Contact your physician immediately if ingested Use this device only with the InTENSity at Home brand electrodes E201 6R POSSIBLE ADVERSE RE...

Page 13: ...13 Instruction Manual E200 6R Unit 2 Lead Wires 4 AAA Batteries 4 2 x 2 Electrodes PACKAGE CONTENTS...

Page 14: ...nsity 4 CH 2 Output Intensity Increase or decrease channel 2 intensity 5 Pause Button Pause resumes treatment 6 Treatment Records To check treatment records and usage history 7 Power Press to turn on...

Page 15: ...or lock 7 Hour Indicator For treatment records only 8 Channel Intensity Graphic display CH1 CH2 9 Channel Intensity Numeric display CH1 CH2 10 Therapeutic Body Part Displays selection 11 Treatment Tim...

Page 16: ...o your federal state and local regulations As a consumer you are obligated by law to properly dispose depleted batteries CAUTION Keep the batteries and TENS unit out of the reach of children Do not di...

Page 17: ...pins NOTE You must use at least 2 electrode pads Step 3 Connect Lead Wires to the Device 1 Make sure the device is turned OFF 2 Holding the insulated portion push the plug end of the lead wire into CH...

Page 18: ...ual This is a powerful piece of equipment and is neither a toy nor a gadget Step 4 Electrode Placement 1 Ensure skin is clean and thoroughly dry before applying the electrodes 2 Remove the electrodes...

Page 19: ...s P1 and P2 You can toggle between them by pressing the body part button repeatedly 3 P1 is for acute pain and P2 is for chronic pain Step 5 Turn the Power On 1 Press the button on the top right side...

Page 20: ...e Width Cycle Time Neck P1 20 Min 80 100Hz 100 150 s 10 Sec P2 20 Min 4Hz 150 200 s 20 Sec Shoulder P1 20 Min 80 100Hz 100 s 10 Sec P2 20 Min 10Hz 220 260 s 20 Sec Hand P1 20 Min 100Hz 100 s Fixed P2...

Page 21: ...idth Cycle Time Low back P1 20 Min 80 100Hz 120 s 10 Sec P2 20 Min 4Hz 200 260 s 20 Sec Knee P1 20 Min 120Hz 100 120 s 10 Sec P2 20 Min 1 10Hz 150 200 s 20 Sec Foot P1 20 Min 80 120Hz 100 120 s 10 Sec...

Page 22: ...crease the output intensity CAUTION 1 If the electrodes are not placed firmly on the skin or the device is not connected to the electrodes and the output intensity level is equal to 10mA or greater th...

Page 23: ...Treatment Timer 1 All 12 presets are set to a 20 minute timer 2 When the timer counts all the way down to zero the device will stop and return to the home screen Step 10 Turn the Power Off 1 Press and...

Page 24: ...Do not place on cuts or damaged skin The InTENSity at Home Dual Channel TENS is for single person use Place the electrode on the protective transparent film when not in use Safety Lock The lock funct...

Page 25: ...e permanently deleted and can not be restored if you complete the deletion process Treatment Record Log The treatment record log can store up to 60 treatments The treatment record log can store up to...

Page 26: ...at least 2 but no more than 6 apart The skin must be clean and dry Press on firmly making good contact between the skin and the electrodes Never remove electrodes from the skin while the device is tu...

Page 27: ...27 Knee Elbow Pain Carpal Tunnel Wrist Pain Lower Back Lumbar Pain ELECTRODE PLACEMENT GUIDE CONT D...

Page 28: ...28 Foot Calf Pain Caution Do not stimulate both feet or calves simultaneously ELECTRODE PLACEMENT GUIDE CONT D...

Page 29: ...er off and remove the lead wires from the electrodes 2 Wash the electrodes when the adhesive surface becomes dirty and or the electrodes are difficult to attach to the skin Washing the Electrode Pads...

Page 30: ...body Never remove the self adhesive electrodes from the skin while the device is still turned on If replacement electrodes are necessary use only electrodes that are the same size 2 x 2 as the electr...

Page 31: ...to the plastic film and then store them and the lead wires in the sealed package Storing the Unit 1 Place the unit electrodes lead wires and manual back into the retail packaging Store the box in a co...

Page 32: ...ditions 10 C 50 C 14 F to 122 F Relative Humidity 10 90 Atmospheric Pressure 700 1060 Hpa Dimensions 12 0 7 0 2 7cm 4 7 x2 8 x1 1 L W H Weight 100g 3 5 oz Without Battery Electrode Detection Function...

Page 33: ...Clean treatment area of dirt and or oily substances lotion Lead wires are old worn damaged Replace with new lead wires Stimulation Is Uncomfortable Intensity is too high Decrease intensity Electrodes...

Page 34: ...Ineffective Improper electrode placement Reposition electrodes Unknown Contact tech support at 1 800 3 ROSCOE 376 7263 Skin Becomes Red And Or You Feel A Stabbing Pain Using the electrode on the same...

Page 35: ...w electrode If problems persist Contact your physician Output Current Stops During Therapy The electrode came off the skin or the edges are lifting away from the skin Turn off the device and reapply t...

Page 36: ...legal obligation Electrical devices are recyclable material and should not be disposed of with household waste after their useful life Help us to protect the environment and save resources and take t...

Page 37: ...ectromagnetic emissions for medical devices Medical devices manufactured for Compass Health Brands conform to this IEC60601 1 2 2007 standard for both immunity and emissions Nevertheless special preca...

Page 38: ...F emissions CISPR 11 Group 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipmen...

Page 39: ...kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 Not appli...

Page 40: ...vice including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80 MHz to 800...

Page 41: ...oadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If t...

Page 42: ...netic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and as recommended below according to the maximum output power of the communic...

Page 43: ...ut power rating of the transmitter in watts W according to the transmitter manufacturer NOTE I At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guideli...

Page 44: ...N DURATION TO A PERIOD OF 120 DAYS FROM THE DATE OF PURCHASE BY THE ORIGINAL PURCHASER In the event that this product is found by Compass Health Brands to not meet the above limited warranty as purcha...

Page 45: ...ENSity at Home product and packaging The original receipt showing date of purchase Detailed description of the problem Serial Number InTENSity at Home Model ____________________________________ Serial...

Page 46: ...46 Middleburg Heights OH 44130 800 376 7263 compasshealthbrands com Made in China Hecho en China Manufactured for...

Reviews: