D-0123672-H – 2022/05
Affinity Compact - Instructions for Use - US
Page
44
3.4 The HIT440 screen
The following section describes the elements of the HIT screen
Menu
provides access to Print, Edit, View, Mode, Setup, and Help.
button
allows you to print only the test results currently
displayed on the screen. To print multiple tests on one page, select
Print then Print Layout
Save & New Session
button saves the current session in Noah or
OtoAccess® and opens a new one.
Save & Exit
button saves the current session in Noah or
OtoAccess® and exits the Suite.
Change Ea
r button allows you to toggle between right and left ear.
Right click on the ear icon to view
both ears
.
Toggle between single and combined screen
button toggles
between viewing one or multiple measurements in the same HIT
graph.
Toggle between single and continuous measurement
button
toggles between running a single sweep or having a test signal
running continuously until pressing STOP.
Freeze curve
allows for taking a snapshot of a HIT curve when
testing with broadband signals. In other words, the curve freezes at
a particular moment while the test continues.
NOTE:
The Freeze Curve option only works in an end-user-created
protocol, for broadband (ex: ISTS) signals in the continuous mode.
List of Protocols
allows you to
select a test protocol (default or
user defined) to use in the current test session.