D-0123672-H – 2022/05
Affinity Compact - Instructions for Use - US
Page
76
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The
AFFINITY COMPACT
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
AFFINITY COMPACT
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be greater
than 30%.
Electrical fast
transient/burst
IEC61000-4-4
+2 kV for power supply
lines
+1 kV for input/output
lines
+2 kV for power
supply lines
+1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
residential environment.
Surge
IEC 61000-4-5
+1 kV differential mode
+2 kV common mode
+1 kV differential
mode
+2 kV common mode
Mains power quality should be
that of a typical commercial or
residential environment.
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
< 5% UT
(>95% dip in UT) for 0.5
cycle
40% UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(>95% dip in UT) for 5
sec
< 5% UT (>95% dip in
UT)
for 0.5 cycle
40% UT (60% dip in
UT) for
5 cycles
70% UT (30% dip in
UT) for
25 cycles
<5% UT
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of the
AFFINITY
COMPACT
requires continued
operation during power mains
interruptions, it is
recommended that the
AFFINITY COMPACT
be
powered from an
uninterruptable power supply
or its battery.
Power frequency
(50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or residential
environment.
Note:
UT is the A.C. mains voltage prior to application of the test level.