D-0128073-C – 2022/05
AT235 - Instructions for Use - US
Page 41
4.3 Cleaning the Probe Tip
Step 4: Assemble the probe again.
Step 3: Replace the cleaned tip.
Step 1: Unscrew the probe cap and remove the probe tip.
Step 2: Thread the stiff end of the cleaning brush into one of the tubes from inside. Pull the cleaning
floss completely through the probe tip tube. Clean each of the three tubes. Discard floss after use.
Notice:
Diagnostic probe Clinical probe
Only insert the cleaning brush from inside out, this will ensure that the dirt is pushed out of
the probe instead of into the probe and as well protect the gasket from being damaged.
Never clean inside the probe holes.
Summary of Contents for AT235
Page 1: ...Instructions for Use US AT235 Science made smarter D 0128073 C 2022 05 ...
Page 2: ......
Page 8: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 4 ...
Page 18: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 14 ...
Page 42: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 38 ...
Page 48: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 44 ...