D-0128073-C – 2022/05
AT235 - Instructions for Use - US
Page 46
Reflex Functions
Signal sources:
Tone - Contra, Reflex:
250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz,
Wide Band, High and Low pass
Tone - Ipsi, Reflex:
500, 1000, 2000, 3000, 4000 Hz wide band, high and
low pass.
NB noise – Contra, Reflex 250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz
NB noise – Ipsi, Reflex
1000, 2000, 3000, 4000 Hz
Stimulus duration:
750 ms
Reflex Acceptance
Adjustable between 2% and 6%, or 0.05 – 0.15 ml
change of ear canal volume.
Intervals
Down to 1 dB step size.
Intensity max
90, 100, 120 dBHL.
Outputs:
Contra Earphone:
TDH39 earphone, DD45 earphone, IP30 single contra
insert and/or IP30insert for Reflex measurements.
Ipsi Earphone:
Probe earphone incorporated in the probe system for
Reflex measurements.
Probe connection
Connection of the electrical and air system to the
probe.
Test types:
Manual Reflex
Manual control of all functions.
Automated Reflex
Single intensities
Reflex growth
Reflex Decay
Automatic, 10 dB above threshold and manually
controlled with stimulus durations of 10.
Reflex latency
Automated, first 300 ms from stimulus start.
Audiometer Functions
Signals:
Frequencies Hz: Intensities dB HL:
125
-10 to 70
250
-10 to 90
500
-10 to 100
1000
-10 to 100
2000
-10 to 100
3000
-10 to 100
4000
-10 to 100
6000
-10 to 100
8000
-10 to 90
Test types
Auto Threshold Determination (Modified Hughson Westlake).
Auto testing: duration 1-2 s adjusted in 0.1 s intervals
Summary of Contents for AT235
Page 1: ...Instructions for Use US AT235 Science made smarter D 0128073 C 2022 05 ...
Page 2: ......
Page 8: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 4 ...
Page 18: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 14 ...
Page 42: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 38 ...
Page 48: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 44 ...