Guidance and manufacturer’s declaration — electromagnetic immunity
The AT235
is intended for use in the electromagnetic environment specified below. The customer or the user
of the AT235 should assure that it is used in such an environment,
Immunity test
IEC / EN 60601
test level
Compliance level
Electromagnetic environment –
guidance
Conducted RF
IEC / EN 61000-4-6
Radiated RF
IEC / EN 61000-4-3
3 Vrms
150kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to any
parts of the AT235, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
P
d
2
,
1
=
P
d
2
,
1
=
80 MHz to 800 MHz
P
d
3
,
2
=
800 MHz to 2,5 GHz
Where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, (a)
should be less than the
compliance level in each
frequency range (b)
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
(a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
AT235
is used exceeds the applicable RF compliance level above, the AT235
should be observed to verify
normal operation, If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the AT235.
(b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Summary of Contents for AT235
Page 1: ...Instructions for Use US AT235 Science made smarter D 0128073 C 2022 05 ...
Page 2: ......
Page 8: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 4 ...
Page 18: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 14 ...
Page 42: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 38 ...
Page 48: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 44 ...