D-0128073-C – 2022/05
AT235 - Instructions for Use - US
Page 56
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is essential to use only
the following accessories:
ITEM
MANUFACTURER
MODEL
Power supply unit UE65
Interacoustics
UE65-240250SPA
Clinical Probe
Interacoustics
Clinical probe system 1077
Diagnostic prob
Interacoustics
Diagnostic probe system 1077
DD45C Contra Headset DD45C
contra headset P3045
Interacoustics
DD45C
DD65 v2 Passive attenuation
audiometric headset
Interacoustics
DD65 v2
IP30 Insert Phone 10ohm single
Interacoustics
IP30C
DD45 Audiometric Headset P3045
Interacoustics
DD45
IP30 Insert Phone 10ohm set
Interacoustics
IP30
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the cable types and cable
lengths are as specified below:
Description
Length
Screened?
Mains Cable
2.0m
Unscreened
USB Cable
2.0m
Screened
Clinical Probe
2.0m
Unscreened
Diagnostic probe
2.0m
Unscreened
DD45C contra headset P3045
0.4m
Screened
DD65 v2 Passive attenuation
audiometric headset
2.0m
Screened
IP30 Insert Phone 10ohm single
contra
0.5m
Screened
DD45 Audiometric Headset P3045
0.5m
Screened
IP30 Insert Phone 10ohm set
0.5m
Screened
Summary of Contents for AT235
Page 1: ...Instructions for Use US AT235 Science made smarter D 0128073 C 2022 05 ...
Page 2: ......
Page 8: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 4 ...
Page 18: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 14 ...
Page 42: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 38 ...
Page 48: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 44 ...