Service Manual
AT235
IX
6.5 Survey of reference and max. hearing level Tympanometer
DD45 Contra
Center
Frequency
Att.
setting
Reading
Tone
NB
[Hz]
[dB HL
[dB HL]
[dB HL]
125
85
75
60
250
85
100
85
500
85
115
100
750
85
120
105
1000
85
120
105
1500
85
120
100
2000
85
115
100
3000
85
125
105
4000
85
115
105
6000
85
110
90
8000
75/70
105
95
WB
85
-
120
LP
85
-
120
HP
85
-
120
Table 10
TDH39 Contra
Center
Frequency
Att.
setting
Reading
Tone
NB
[Hz]
[dB HL
[dB HL]
[dB HL]
125
85
80
65
250
85
100
85
500
85
115
100
750
85
120
105
1000
85
120
105
1500
85
120
105
2000
85
120
105
3000
85
120
105
4000
85
120
105
6000
85
120
100
8000
75/70
105
95
WB
85
-
115
LP
85
-
120
HP
85
-
115
Table 11
IOW IPSI Sp.
Center
Frequency
Att.
Setting
Reading
Tone
NB
[Hz]
[dB HL
[dB HL]
[dB HL]
125
85
70
60
250
85
85
75
500
85
100
85
750
85
100
85
1000
85
105
90
1500
85
110
90
2000
85
105
90
3000
85
95
90
4000
85
100
85
6000
85
85
80
8000
75/70
80
75
WB
85
-
95
LP
85
-
100
HP
85
-
95
Table 12
IP30 Contra / EAR 3A Contra
Center
Frequency
Att.
setting
Reading
Tone
NB
[Hz]
[dB HL
[dB HL]
[dB HL]
125
85
100
85
250
85
110
100
500
85
115
105
750
85
120
110
1000
85
120
110
1500
85
120
110
2000
85
120
110
3000
85
120
110
4000
85
120
105
6000
85
105
100
8000
75/70
90
85
WB
85
-
115
LP
85
-
120
HP
85
-
120
Table 13
CIR 33 Contra
Center
Frequency
Att.
setting
Reading
Tone
NB
[Hz]
[dB HL
[dB HL]
[dB HL]
125
85
90
70
250
85
100
85
500
85
110
100
750
85
110
105
1000
85
115
105
1500
85
115
105
2000
85
115
105
3000
85
115
105
4000
85
110
100
6000
85
95
95
8000
80/85
75
80
WB
85
-
115
LP
85
-
115
HP
85
-
115
Table 14