Service Manual
AT235
6
2.3 Disassembling the Cabinet
To disassemble the instrument, the top must be
separated from the bottom cabinet. The
cabinets are separated by removing the
indicated screws with a Torx T10.
The panels are now loose for dismantling the parts.
Be careful whilst separating as the parts are still
attached with cables on the inside. Please note that
the screen is not fully attached anymore.
By tilting the top, the cabinet is open for service.
Carefully, free the ribbon cable from the cabinet.
After releasing the ribbon cable and the display
cable, the bottom and top cabinets are separated.
Fig. 16
Fig. 12
Fig. 13
Fig. 14
Fig. 15