Technical Specifications
General
Medical CE-mark:
The CE-mark in combination with MD symbol indicates that Interacoustics A/S
meets the requirements of the Medical Device Regulation (EU) 2017/745. Approval
of the quality system is made by TÜV – identification no. 0123.
Standards:
Safety:
IEC 60601-1, Class I, Type B applied parts
EMC:
IEC 60601-1-2
Impedance:
IEC 60645-5 (2004)/ANSI S3.39 (2012), Type 1
Audiometer:
IEC60645-1 (2012)/ANSI S3.6 (2010), Type 4
Operation
environment:
Temperature:
15 – 35
°
C
Relative Humidity:
30 – 90%
Ambient Pressure:
98kPa – 104kPa
Warm-up Time:
1 minute
Display
10 inch high resolution color display 1024x600
Transport & Storage: Storage Temperature:
Transport Temperature:
Rel. Humidity:
0
°
C – 50
°
C
-20 – 50
°
C
10 – 95%
Internal storage
500 clients and 50.000 sessions
Internal Battery
CR2032 3V, 230mAh, Li. Not serviceable by user.
PC control:
USB:
Input/output for computer communication.
Data can be sent to and saved on the PC and stored
in OtoAccess® (Diagnostic Suite sync module is
required).
Thermal printer
(Optional):
Type: MPT-III
Thermal MPT-III/HM-E300 printer with recording paper
in rolls. Print on command via USB
Power supply
UES65-240250SPA3
Use only specified power supply unit model UES65-
240250SPA3
Input: 100-240VAC 50/60Hz, 2.0 A
Output: 24.0 VDC, 2.5A
Dimensions
H x W x L
29 x 38 x 7,5 cm
AT235 Weight
2.5 kg
Impedance Measuring System
Probe tone:
Frequency:
Level:
226 Hz, 678 Hz, 800 Hz, 1000 Hz; pure tones; ±1%
85 dB SPL (≈ 69 dB HL) ±1.5 dB
Air pressure:
Control:
Indicator:
Range:
Pressure limitation:
Pump Speed:
Automatic.
Measured value is displayed on the graphical display.
-600 to +400 daPa. ±5%
-750 daPa and +550 daPa.
Automatic, Fast 300 daPa/s, Medium 200 daPa/s,
Slow 100 daPa/s, Very slow 50 daPa/s.
Compliance:
Range:
0.1 to 8.0 ml at 226 Hz probe tone (Ear volume: 0.1 to
8.0 ml) and 0.1 to 15 mmho at 678, 800 and 1000 Hz
probe tone. All ±5%
Test types:
Tympanometry
Automatic, where the start and stop pressure can be
user-programmed in the setup function.
Manual control of all functions.
Eustachian tube function 1
- Non perforated eardrum
Williams test
Eustachain tube function 2
- Perforated eardrum
Toynbee test
Eustachian tube function 3
- Patulous Eustachian tube
Continuous sensitive impedance measurement
Summary of Contents for AT235
Page 1: ...Instructions for Use US AT235 Science made smarter D 0128073 C 2022 05 ...
Page 2: ......
Page 8: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 4 ...
Page 18: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 14 ...
Page 42: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 38 ...
Page 48: ...D 0128073 C 2022 05 AT235 Instructions for Use US Page 44 ...