8104738 Rev. 2 – 2019/03 

 

Page 

1

 of 

2

 

Quick Guide 

Titan ABRIS handheld  

 

PREPARE THE EQUIPMENT 

1.  Turn on Titan by pressing the R or L button. 

2.  Choose the test protocol from the list using the up & down arrow keys and then press 

Select.

 

3.  To use a different protocol than the one selected, press 

Protocol

 and then repeat step 2 above. 

4.  Select the transducer (only if more than one connected to the preamplifier). 

5.  Select the ear for testing. 

 

TEST ENVIRONMENT

 

The ideal test environment is a quiet room where lights and other electronic equipment are turned off. 

 

PREPARE THE INFANT  

Patient state 

 

The infant should be sleeping or in quiet relaxed state.    

Sucking, blinking, crying or movement may affect testing. 

 

Skin preparation 

Typically, no skin preparation is required.   

If the infant’s skin is oily, the electrode placement sites should be cleaned using 

alcohol wipes and/or skin preparation gel. 

Use a conductive gel with the electrodes to improve impedances. 

 

Place electrodes 

 

Place surface electrodes using the desired montage (mastoid or nape). 

    

 

 

Connect cables  

 

Connect the cables from the PreAmplifier to the respective surface electrodes. 

 

CHECK IMPEDANCE 

Impedances are indicated by the green/amber dots on the infant as well as in numerical format on screen. 

 

When the dot is amber, this means the impedance 

is poor (> 40kΩ). In this instance, it may be 

necessary to clean the skin and replace the 

surface electrode.   

 

 

NB.  Testing is possible when impedances are 

poor.  This may, however, affect test time and 

measurements. 

 