15. Adjust

 

the

 

injector

 

by

 

performing

 

one

 

of

 

the

 

following:

Bottom

 

of

 

Floating

 

Ball

 

Reads

 

Greater

 

Than

 

35

 ‐ 

Less

 

disinfectant

 

needs

 

to

 

be

 

added

 

to

 

the

 

mixture.

 

Loosen

 

the

 

screw

 

on

 

the

 

injector

 

to

 

draw

 

less

 

disinfectant

 

into

 

the

 

solution.

Bottom

 

of

 

Floating

 

Ball

 

Reads

 

Less

 

Than

 

35

 ‐ 

More

 

disinfectant

 

needs

 

to

 

be

 

added

 

to

 

the

 

mixture.

 

Tighten

 

the

 

screw

 

on

 

the

 

injector

 

to

 

draw

 

more

 

disinfectant

 

into

 

the

 

solution.

16. When

 

the

 

bottom

 

of

 

the

 

floating

 

ball

 

is

 

near

 

35,

 

turn

 

the

 

black

 

adjustment

 

knob

 

on

 

the

 

flow

 

meter

 

until

 

the

 

bottom

 

of

 

the

 

floating

 

ball

 

is

 

at

 

35.

17. Place

 

the

 

bucket/pail

 

under

 

the

 

burp

 

fitting

 

to

 

collect

 

the

 

disinfectant.

 

Use

 

a

 

funnel

 

if

 

necessary.

 

18. Note

 

the

 

number

 

of

 

ounces

 

in

 

the

 

graduated

 

plastic

 

bottle.

19. Let

 

the

 

bucket/pail

 

fill

 

to

 

one

 

gallon.

20. Turn

 

the

 

disinfectant

 

valve

 

to

 

the

 

off

 

position.

21. Remove

 

the

 

disinfectant

 

draw

 

tube

 

from

 

the

 

plastic

 

bottle.

22. Examine

 

the

 

amount

 

of

 

disinfectant

 

remaining

 

in

 

the

 

plastic

 

bottle.

 

23. Subtract

 

the

 

amount

 

of

 

disinfectant

 

remaining

 

from

 

the

 

amount

 

noted

 

in

 

STEP

 

18.

 

24. Perform

 

one

 

of

 

the

 

following:

 

Approximately

 

½

ounce

 

of

 

disinfectant

 

has

 

been

 

used

 ‐ 

Proceed

 

to

 

STEP

 

23.

Much

 

more

 

or

 

much

 

less

 

than

 

½

ounce

 

of

 

disinfectant

 

has

 

been

 

used

 ‐ 

Repeat

 

STEPS

 

1

18.

25. Position

 

the

 

disinfectant

 

draw

 

tube

 

into

 

the

 

disinfectant

 

solution.

26. Disinfect

 

the

 

tub.

 

Refer

 

to

 

the

 

tub

 

owner’s

 

manual

 

or

 

to

 

the

 

Closed

 

Loop

 

Disinfecting

 

Procedure

 

instruction

 

sheet,

 

part

 

number

 

1154273.

FIGURE 1   

Calibrating the Disinfectant Flowmeter

Service Access 

Cabinet

Floating 

Ball

Adjustment 

Knob

Flow 

Meter

Disinfectant Valve

DETAIL “A” - 

FLOW METER

Screw

Injector

Disinfectant Siphon Tube

Invacare Continuing Care Group

www.invacare-ccg.com

USA

One Invacare Way

Elyria, Ohio USA

44036-2125

Canada

994 Hargrieve Rd.

London Ontario

N6E IP5 Canada

 

All rights reserved. Trademarks are identified by the symbols ™ and ®. All 

trademarks are owned by or licensed to Invacare Corporation unless 

otherwise noted.

Betadine is a registered trademark of Purdue Pharma L.P.

© 2008 Invacare Corporation

Part No 1154272

For Customer Service call: 800-668-2337

Rev A - 04/08

Reviews: