Invacare® Homecare Beds
4.5 Assembling the Spring Fabric
1.
Lift the head spring
C
away from the bed frame to give
slack to the fabric.
2.
Connect the spring fabric
A
together with the links
B
provided.
4.6
Adjusting and Reconnecting the Head
Section Pull Tube
1.
To disconnect the head pull tube assembly
A
from the
side frame of the foot spring, remove the hitch pin
E
and grommet/washer
F
from the clevis pin
G
.
2.
Extend the inner pull tube shaft
C
manually until the
spring button "clicks" into the adjustment hole of the
outer pull tube
D
.
3.
Connect the pull tube end assembly
A
to the lift arm
of the head spring as follows:
a. Insert the clevis pin
G
into the lift arm slot
H
.
b. Replace the grommet/washer
F
.
c. Secure with hitch pin
E
.
22
1114836-H-05