SECTION 2—FRONT HANGERS
1 0
S E C T I O N 2 — F R O N T H A N G E R S
W A R N I N G
After ANY adjustments, repair or service and before use, make sure all
attaching hardware is tightened securely - otherwise injury or damage
may occur.
A 50mm clearance between the bottom of the footplate and the ground/floor
MUST be maintained at all times.
Adjusting Footrest Height
NOTE: For this procedure, refer to FIGURE 2.1.
1. Remove calf strap, if so equipped.
2. Unscrew locking screw.
3. Position the footrest assembly to the desired height.
NOTE: A 5 cm clearance between the bottom of the footplate and the ground/floor MUST
be maintained at all times.
4. If the distance between the bottom of footplate and the ground/floor is not 50 mm
minimum, adjust footrest tube accordingly.
5. Using the locking screw, secure the footrest tube to the footrest support. Securely
tighten.
6. If necessary, repeat STEPS 2
-
5 to adjust remaining footrest.
7. Reinstall the calf strap, if so equipped.
5 cm mini
Locking screw
FIGURE 2.1
Adjusting Footrest Height
Assembly description
Screw
Tool
Torque
Hanger pivot
H FL CS-M 6x30
Allen wrench #4
5 ±1 Nm
Lock-housing mechanism
Socket button cap
ST4,8x16
Torx T25
5 ±1 Nm
Footplate tube
H-M 6x25
Flat hexa wrench #10
5 ±1 Nm
Footplate axle
Socket Cap-M 6x55
Allen wrench #5
4 ±1 Nm