Power frequency (50/60 Hz)
magnetic field IEC 61000-4-
8
3 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The AVIVA FX Power Wheelchair is intended for use in the electromagnetic environment specified below. The customer or the user of the AVIVA
FX Power Wheelchair should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 TEST
LEVEL
Compliance level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Portable and mobile RF communications equipment should be
used no closer to any part of the AVIVA FX Power Wheelchair,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter.
Radiated RF IEC
61000-4-3
3 V/m
80 MHz to 2,5 GHz
20 V/m
26 MHz to 2.5 GHz per
ISO 7176-21
Recommended separation distance
(3.5/ 3 V/m = 1.667)
26 MHz to 800 MHz (3.5/ 20 V/m = 0.175)
800 MHz to 2,5 GHz (7/ 20V/m = 0.35)
60101916-A
19
3 Electromagnetic Compatibility (EMC) Information