Invacare® Modulite
Installing Holder
1.
Install parts in reverse order. Tighten screw
B
to
6 Nm.
2.
Test all functions.
5.8.2 Replacing remote bracket for attendant
Removing
•
3 mm Allen key
•
4 mm Allen key
1.
Loosen screws
C
and nuts
A
.
2.
Remove bracket
B
.
Installing
1.
Mount bracket to tube in desired position.
2.
Tighten screws
C
and nuts
A
.
5.8.3 Replacing Remote Bracket for
DLX-ACU200
Removing
•
3 mm Allen key
•
5 mm Allen key
•
10 mm wrench
1.
Loosen screws
D
. Remove remote
A
from clamp
B
.
2.
Remove screws
C
and spacers
B
and clamp
A
from
gripper clamp
D
.
3.
26