9 Technical data
9.1 Dimensions
All measurements are stated in cm. All angles are stated in degrees.
All measurements and angles are stated without tolerances.
Invacare®
reserves the right to change the stated measurements and
angles.
A
B
D
C
Etude Plus low
Etude Plus
Low
High
Low
High
A
219
219
219
219
B
102
102
102
102
C
20–60
27–67
33–73
40–80
D
138–178
145–185
158–198
165–205
The angle between the lower leg section and horizontal is adjustable
from 0º to 15º.
Bed with tilt function
9.2 Weights
Max. patient weight
(provided that the weight of the mattress
and the accessories do not exceed 35 kg)
165 kg
Max. safe working load
(p accessories)
200 kg
Etude Plus standard bed, complete, excluding
accessories
77.3 kg
Etude Plus auto-regression, complete, excluding
accessories
82.7 kg
Etude Plus bed end with extra plate – 1 piece
20.6 kg
Etude Plus bed end without extra plate – 1 piece
17.9 kg
Mattress support, standard head end
24.1 kg
Mattress support with auto-regression, head end
27.9 kg
Mattress support with auto-regression, head end,
without actuator and control box
24.1 kg
Mattress support, foot end
17.4 kg
Scala Basic 2 side rail – 1 piece
7.4 kg
Lifting pole
4.2 kg
Mattress support extension (10 cm)
3.2 kg
Mattress support extension (20 cm)
5.5 kg
9.3 Environmental conditions
Storage
Operation and
Transportation
Temperature
-10°C to +50°C
+5°C to 40°C
Relative humidity
20% to 75%
Atmospheric
pressure
800 hPa to 1060 hPa
Be aware that when a bed has been stored under low
temperatures, it must be adjusted to operating conditions
before use.
9.4 Electrical system
Voltage supply: Uin = 230 Volt, AC, 50 Hz (AC = Alternating
current)
Maximum current input: Iin max. = 2 A
Intermittent (periodic motor operation): Int = max. 10%, 2 min /
18 min
Degree of protection: IPx6* or IPx4** (depending on configuration)
The control unit, external power supply and motors are protected
according to IPx6. When the bed is equipped with a HL80 hand
control, it is protected according to IPx4
Insulation class: Class II
Type B Applied Part
Applied Part complying with the specified requirements for
protection against electrical shock according to IEC60601-1.
(An applied parts is a part of the medical equipment which is
designed to come into physical contact with the patient or parts that
are likely to be brought into contact with the patient.)
Sound level: 45 to 50 dB (A)
* IPX6 classification means that the electrical system is protected
against water projected in powerful jets from any direction.
** IPX4 classification means that the electrical system is protected
against water splashed against the component from any direction.
The bed is not provided with a mains switch, so the mains
plug is the only separation from the mains.
1567454-A
15
Summary of Contents for Etude Plus
Page 18: ...Notes...
Page 19: ...Notes...