6 Maintenance

6.1 Maintenance schedule

Important

Service and maintenance of the bed must only be performed by personnel who have received the necessary instruction or training.

With normal daily use, service must be carried out according to the service schedule after two years of use and thereafter every second year.

When relocating the bed, service must be carried out according to the service schedule.

Before Use

Ensure that all manual and electrical parts are functioning correctly and are in a secure state.

After three month

Ensure that all manual and electrical parts are functioning, and tighten bolts, screws, nuts, etc.

Every two years

Service should be performed according to the maintenance chart. See below.

Bed / Id. no.: ______________________________________

Date / Initials

Visual inspection of all parts of the bed (plastic deformation and/or wear and tear of
welded joints)

Control all centres of rotation ( mattress support parts)

Check that all motors are running without failures (with regular speed and at low noise)

Check the Rastofix fittings and function

Visual inspection of all cabinets for damages.

Check that mains cable and plug are intact. Check the running of the cable.

Check all remaining cables for damages. Also check the running of the cables.

Check of siderails´ fixing and locking/movement.

Check the wheels (security, braking and free rolling)

Lubrication plan

We recommend lubricating the bed according to the following instructions:

Lubricate with medically clean oil, e.g. KEN-WO 50:

Points of rotation in mattress support and base frame - lubricate with oil.

All of the motors’ tension rod - lubricate with oil.

All motor bearings and Rastofix fitting - lubricate with oil.

The wooden side rails gliding system must not be lubricated with oil - otherwise the wooden bars will move sluggishly.

6.2 Cleaning

Important

Make sure the power plug is removed from the socket outlet before cleaning.

All plugs must be installed during washing of the bed.

The bed should be washed down using a wet sponge, cloth or brush.

The bed does not tolerate cleaning in automatic washing plants or the use of water-jet based cleaning equipment.

Use ordinary household cleaning agents. Never use acids, alkalines or solvents such as acetone or cellulose thinner.

The hand control, motors and control unit may be washed with brush and water, but not with pressurized water.

Dry the bed after cleaning.

12

1567454-A

Summary of Contents for Etude Plus

Page 1: ...Invacare Etude Plus Etude Plus Etude Plus Low EN Bed User Manual This manual MUST be given to the user of the product BEFORE using this product read this manual and save for future reference...

Page 2: ...dentified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted This bed has been tested and approved according to IEC 60601 2 52 Beds with...

Page 3: ...Emergency release of the backrest or thigh leg section 8 5 Accessories 9 5 1 List of available accessories 9 5 2 Side rails 9 Steel side rails 9 Line and Britt V side rails 9 5 3 Lifting pole 10 5 4...

Page 4: ...ined to be used according to Application environment 3 Long term care in a medical area where medical supervision is required and monitoring is provided if necessary and medical electrical equipment u...

Page 5: ...p through the openings between the side rails or through the opening between the side rail and the mattress support Pay special attention if the bed is used for the care of patients with small body di...

Page 6: ...0000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 A B C D E F G H I J Scope of delivery A Mattress support head end B 4...

Page 7: ...four hooks at the head end 8 Fasten the safety clip to the control box 3 4 Dismantle the bed 1 Dismantle side rails and lifting pole 2 Bring the bed to its lowest position and adjust all mattress sup...

Page 8: ...the bed is positioned correctly at least one castor at the head end and one castor at the foot end must be braked WARNING Risk of injury The bed s castors must be locked whilst nursing a patient in be...

Page 9: ...d bed end When using steel side rails always make sure that the distance between the bed end and the side rail s handle upper edge is less than 6 cm in the head end and more than 32 cm in the foot end...

Page 10: ...ommended to use a mattress support extension Fitting the mattress support extension 1 Disassemble the bed and remove the inserts 2 Mount the mattress support extension between both parts of the mattre...

Page 11: ...Accessories both Line side rails standard and extendable Min density 38kg m3 5 6 Shipping brackets Transport brackets 1 2 3 4 1567454 A 11...

Page 12: ...ll cabinets for damages Check that mains cable and plug are intact Check the running of the cable Check all remaining cables for damages Also check the running of the cables Check of siderails fixing...

Page 13: ...out wheelie bin symbol is placed on this product to encourage you to recycle wherever possible Please be environmentally responsible and recycle this product through your recycling facility at its end...

Page 14: ...is not running The relay in the control unit makes a clicking noise Control unit is defective Replace the control unit Control unit is defective Replace the control unit Mains indicator lights up but...

Page 15: ...cm 5 5 kg 9 3 Environmental conditions Storage Operation and Transportation Temperature 10 C to 50 C 5 C to 40 C Relative humidity 20 to 75 Atmospheric pressure 800 hPa to 1060 hPa Be aware that when...

Page 16: ...tact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrostatic transient bur...

Page 17: ...observed additional measures may be necessary such as reorienting or relocating the medical bed b Over the frequency range 150 kHz to 80 MHz field strengths should be less than V1 V m At 80 MHz and 80...

Page 18: ...Notes...

Page 19: ...Notes...

Page 20: ...0 7085 ireland invacare com www invacare ie United Kingdom Invacare Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ Tel 44 0 1656 776 222 Fax 44 0 1656 776 220 uk invacare com www invacare c...

Reviews: