8 Troubleshooting

8.1 Troubleshooting electrical system

Symptom

Possible cause

Remedy

Mains is not connected

Connect mains

Fuse in the control unit is blown

Replace the control unit

Mains indicator does not light up

Control unit is defective

Replace the control unit

Motor plug is not fully inserted into the
control unit.

Insert the motor plug properly into the
control unit

The motor is defective.

Replace the motor

Motor cable is damaged.

Replace the cable

Mains indicator lights up, but the motor is not
running. The relay in the control unit makes a
clicking noise.

Control unit is defective

Replace the control unit

Control unit is defective

Replace the control unit

Mains indicator lights up, but the motor is not
running. No relay sound is heard from the
control unit.

Hand control is defective

Replace the hand control

Control unit is defective

Replace the control unit

Control unit is in order except for one
direction on one channel.

Hand control is defective

Replace the hand control

Motor is running, but the piston rod does not
move.

The motor cannot lift full load.

Motor noise, but no movement of piston rod.

Piston rod operates inwards and not outwards.

Motor is damaged

Replace the motor

14

1567454-A

Summary of Contents for Etude Plus

Page 1: ...Invacare Etude Plus Etude Plus Etude Plus Low EN Bed User Manual This manual MUST be given to the user of the product BEFORE using this product read this manual and save for future reference...

Page 2: ...dentified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted This bed has been tested and approved according to IEC 60601 2 52 Beds with...

Page 3: ...Emergency release of the backrest or thigh leg section 8 5 Accessories 9 5 1 List of available accessories 9 5 2 Side rails 9 Steel side rails 9 Line and Britt V side rails 9 5 3 Lifting pole 10 5 4...

Page 4: ...ined to be used according to Application environment 3 Long term care in a medical area where medical supervision is required and monitoring is provided if necessary and medical electrical equipment u...

Page 5: ...p through the openings between the side rails or through the opening between the side rail and the mattress support Pay special attention if the bed is used for the care of patients with small body di...

Page 6: ...0000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 00000000 A B C D E F G H I J Scope of delivery A Mattress support head end B 4...

Page 7: ...four hooks at the head end 8 Fasten the safety clip to the control box 3 4 Dismantle the bed 1 Dismantle side rails and lifting pole 2 Bring the bed to its lowest position and adjust all mattress sup...

Page 8: ...the bed is positioned correctly at least one castor at the head end and one castor at the foot end must be braked WARNING Risk of injury The bed s castors must be locked whilst nursing a patient in be...

Page 9: ...d bed end When using steel side rails always make sure that the distance between the bed end and the side rail s handle upper edge is less than 6 cm in the head end and more than 32 cm in the foot end...

Page 10: ...ommended to use a mattress support extension Fitting the mattress support extension 1 Disassemble the bed and remove the inserts 2 Mount the mattress support extension between both parts of the mattre...

Page 11: ...Accessories both Line side rails standard and extendable Min density 38kg m3 5 6 Shipping brackets Transport brackets 1 2 3 4 1567454 A 11...

Page 12: ...ll cabinets for damages Check that mains cable and plug are intact Check the running of the cable Check all remaining cables for damages Also check the running of the cables Check of siderails fixing...

Page 13: ...out wheelie bin symbol is placed on this product to encourage you to recycle wherever possible Please be environmentally responsible and recycle this product through your recycling facility at its end...

Page 14: ...is not running The relay in the control unit makes a clicking noise Control unit is defective Replace the control unit Control unit is defective Replace the control unit Mains indicator lights up but...

Page 15: ...cm 5 5 kg 9 3 Environmental conditions Storage Operation and Transportation Temperature 10 C to 50 C 5 C to 40 C Relative humidity 20 to 75 Atmospheric pressure 800 hPa to 1060 hPa Be aware that when...

Page 16: ...tact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrostatic transient bur...

Page 17: ...observed additional measures may be necessary such as reorienting or relocating the medical bed b Over the frequency range 150 kHz to 80 MHz field strengths should be less than V1 V m At 80 MHz and 80...

Page 18: ...Notes...

Page 19: ...Notes...

Page 20: ...0 7085 ireland invacare com www invacare ie United Kingdom Invacare Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ Tel 44 0 1656 776 222 Fax 44 0 1656 776 220 uk invacare com www invacare c...

Reviews: