Chapter 5 : Diagnostics
19
5 Diagnostics
SHARK is not user serviceable. Specialized tools are necessary for the
repair of any SHARK component.
5.1 Introduction
An abnormal condition is indicated by a flash code on the Shark Control Unit’s
“Information Gauge” (generally the same as the battery gauge). A Flash Code
is a sequence of flashes, separated by a pause, followed by a repetition of the
sequence. Depending on the severity of the condition, the powerchair may or
may not allow driving. In some cases the chair may be allowed to drive but in
a reduced speed (‘limp’) mode.
Note that joystick OONAPU (Out Of Neutral At Power Up) is not a fault.
Simply by removing your hand from the joystick and allowing it to return
to the neutral position, the fault will immediately clear.
If the condition persists after removing your hand, the joystick may be
damaged. Consult a service agent.
Flash Description
1
User Fault
Possible stall timeout or user error
2
Battery Fault
Check the batteries & cabling
3
Left Motor Fault
Check the left motor connections & cabling
4
Right Motor Fault
Check the right motor connections & cabling
5
Left Park Brake Fault
Check the left park brake connections & cabling
6
Right Park Brake Fault
Check the right park brake connections & cabling
7
SHARK Control Unit Fault
Check the SHARK Communications Bus
8
SHARK Power Module Fault
Check the SHARK connections & wiring
9
SHARK Communications Fault
Check the SHARK connections & wiring
10
Unknown Fault
Check all connections & wiring
11
Incompatible Control Unit
Wrong type of Control Unit Connected
Summary of Contents for Hydra C650PW
Page 29: ...29 ...
Page 34: ...7 7 Electromagnetic Compatibility EMC 48 7 8 Contact Details 49 ...
Page 36: ...GBK80260 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 48: ...GBK80260 Issue 1 1 07 2004 14 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 64: ...GBK80260 Issue 1 1 07 2004 30 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 66: ...GBK80260 Issue 1 1 07 2004 32 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 71: ...Chapter 7 Appendices 37 Dynamic SHARK Looms For DK REMA DK PMA Power Module ...
Page 84: ...Installation Manual DK PMA SHARK Power Module by GBK80262 Issue 1 June 2004 ...
Page 87: ......
Page 89: ...GBK80262 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 91: ...GBK80262 Issue 1 1 07 2004 4 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 113: ...GBK80262 Issue 1 1 07 2004 26 Dynamic SHARK DK PMA Looms ...
Page 116: ...Chapter 7 Appendices 29 Dynamic SHARK Connector Kits and Adapters ...