Chapter 5 : Diagnostics
29
5.3
Flash Codes
Flash codes indicate the nature of an abnormal
condition directly from the SHARK Information
Gauge. Without the use of any servicing tools, the
condition can be simply diagnosed.
Flash Code
Description
1
User Fault
Possible stall timeout or user error.
Release the joystick to neutral and try again.
2
Battery Fault
Check the batteries and cabling.
Try charging the batteries.
Batteries may require replacing.
3
Left Motor Fault
Check the left motor, connections and
cabling.
4
Right Motor Fault
Check the right motor, connections and
cabling.
5
Left Park Brake Fault
Check the left park brake, connections and
cabling.
6
Right Park Brake Fault
Check the right park brake, connections and
cabling.
7
SHARK Control Unit
Fault
Check the SHARK Communications Bus
connections and wiring.
Replace the Control Unit.
8
SHARK Power Module
Fault
Check SHARK connections and wiring.
Replace the Power Module.
9
SHARK
Communications Fault
Check SHARK connections and wiring.
Replace the SHARK Control Unit.
10
Unknown Fault
Check all connections and wiring.
Consult a service agent.
11
Incompatible Control
Unit
Wrong type of Control Unit connected.
Ensure the branding of the Power Module
matches that of the Control Unit.
Summary of Contents for Hydra C650PW
Page 29: ...29 ...
Page 34: ...7 7 Electromagnetic Compatibility EMC 48 7 8 Contact Details 49 ...
Page 36: ...GBK80260 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 48: ...GBK80260 Issue 1 1 07 2004 14 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 64: ...GBK80260 Issue 1 1 07 2004 30 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 66: ...GBK80260 Issue 1 1 07 2004 32 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 71: ...Chapter 7 Appendices 37 Dynamic SHARK Looms For DK REMA DK PMA Power Module ...
Page 84: ...Installation Manual DK PMA SHARK Power Module by GBK80262 Issue 1 June 2004 ...
Page 87: ......
Page 89: ...GBK80262 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 91: ...GBK80262 Issue 1 1 07 2004 4 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 113: ...GBK80262 Issue 1 1 07 2004 26 Dynamic SHARK DK PMA Looms ...
Page 116: ...Chapter 7 Appendices 29 Dynamic SHARK Connector Kits and Adapters ...