GBK80262 : Issue 1 – 1/07/2004
10
3.2.4
Motor and Park Brake Connections
The SHARK Power Module has two motor connectors – M1 and M2. These
are typically Right and Left motor, respectively (see note 4 below). Each motor
connector has two motor pins (Positive and Negative), as well as two Park
Brake pins (Positive and Negative). The motor connectors are ‘keyed’ so that
they cannot be swapped or inserted incorrectly.
These notes are in addition to the “General Wiring Notes and
Recommendations” described in Section 3.2.1.
1. It is preferred that the left and right motor harnesses – M1 and M2 -
are of equal length (but also see note 4).
2. For a 60 Amp controller the recommended MINIMUM motor wire
size is 3 mm
2
for runs of up to 400 mm, increasing by 0.5 mm
2
for
each additional 200 mm run length. The heavier the wire, the better
chair performance will be.
3. The length and gauge of wire effects the wire resistance and hence
the optimum Load Compensation setting. Ensure the “Load
Compensation” parameter is tuned to match the chair wiring.
4. Left and right motors must not be physically interchangeable. The
preferred method to ensure this is to use the polarised motor
connectors (Left Motor Connector Housing = GSM61191P, Right
Motor Connector Housing = GSM61192P). However, an alternative
method to guarantee non-transposable motor wiring can be
achieved by providing the exact wire lengths (non equal) required
for the left and right motors.
If necessary, the motor connections can be swapped when
programming SHARK. For this reason, the connectors are not labeled
Left and Right, but M1 and M2, as shown.
Summary of Contents for Hydra C650PW
Page 29: ...29 ...
Page 34: ...7 7 Electromagnetic Compatibility EMC 48 7 8 Contact Details 49 ...
Page 36: ...GBK80260 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 48: ...GBK80260 Issue 1 1 07 2004 14 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 64: ...GBK80260 Issue 1 1 07 2004 30 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 66: ...GBK80260 Issue 1 1 07 2004 32 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 71: ...Chapter 7 Appendices 37 Dynamic SHARK Looms For DK REMA DK PMA Power Module ...
Page 84: ...Installation Manual DK PMA SHARK Power Module by GBK80262 Issue 1 June 2004 ...
Page 87: ......
Page 89: ...GBK80262 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 91: ...GBK80262 Issue 1 1 07 2004 4 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Page 113: ...GBK80262 Issue 1 1 07 2004 26 Dynamic SHARK DK PMA Looms ...
Page 116: ...Chapter 7 Appendices 29 Dynamic SHARK Connector Kits and Adapters ...