Maintenance
8.13 Maintaining the Base Adjustment
The base
A
adjustment should not require any attention other than
adjusting the linkage rods.
CAUTION!
Damage to the lift may occur.
–
The leg actuator must be in the fully closed position
before adjusting the linkage rods.
1.
Close the legs all the way. Refer to Closing/Opening Manual
Legs, page 19 or Closing/Opening Electric Legs, page 20.
2.
Check the squareness of the legs
B
when in the closed position.
3.
Place a square
C
on the inside of the legs and base to determine
the 90° alignment
D
.
4.
Adjust the linkage rods
E
until 90° alignment is achieved.
CAUTION!
Damage to the lift may occur.
–
DO NOT adjust the linkage rods so the legs are at less
than 90° alignment.