Eskal(L839)_Liber(L803)
1.
Perform one of the following:
a. Place the Alternating Pressure Pump on a flat surface
near the bed.
b. Open the hooks
B
on the back of the Alternating
Pressure Pump
A
and install on the outside of the
bed frame
C
.
2.
Place the mattress overlay on the mattress. The air
chamber openings should be at the foot end of the bed.
3.
Slide the two end flaps under the mattress at the head
and foot end of the bed.
4.
Connect air hoses with fittings
F
to the air chamber
openings
D
on the mattress overlay
E
. Check that the
hoses are not kinked or caught under the mattress.
5 Usage
5.1 Safety Information
WARNING!
Insufficient repositioning of the patient can lead to
tissue compression and potential injury formation.
To relieve pressure, it is very important for
the patients to reposition themselves, or to be
repositioned, on a regular basis.
– Before using the product, always consult a
qualified healthcare professional for clinical
judgement.
– Monitor the patient frequently.
CAUTION!
Risk of damage to the device
If holes are present in the device, there is a risk
that liquids may ingress and contamination may
occur.
– Ensure that the device is not jammed or
damaged by sharp edges.
– Do not place hypodermic needles, venflons,
scalpels or other similarly sharp objects on or
under the device.
– Do not use electrically heated blankets directly
on or under the device.
– Ensure that all venflons are taped down
correctly with no sharp edges exposed.
– When using bridging boards or other patient
transfer aids, check for any sharp edges or burrs
before use.
– When using the device on a profiling bed ensure
that the knee break is used before the backrest.
– Attach medical equipment including infusion
pumps and monitors to appropriate bed
accessories.
– Avoid cigarette burns and pet claws puncturing
the device.
5.2 Using the System
A
B
1.
Plug power cord into outlet.
2.
Turn power switch
B
to the ON position. The green light
will be illuminated when the power is on.
3.
Turn knob
A
to maximum setting (clockwise).
Wait approx. 15 minutes for the mattress overlay to
inflate.
Adjusting the Pump
The pump will cycle every five minutes, inflating and deflating
the two air chambers. The pressure of the pump can be
adjusted to provide maximum effectiveness and pressure
relief for the patient.
1.
Ensure the patient is lying on the mattress overlay.
2.
Adjust the pressure to the desired setting. Turn
the knob clockwise to increase pressure or turn in
counter-clockwise to decrease pressure.
3.
Slide the flat portion of your hand between the mattress
and the mattress overlay. There should be a cushion
of air.
The following table provides approximate pressure settings
according to the weight of the patient:
Pressure
level
Weight
Pressure
level
Weight
1 (MIN)
30 kg
6
80 kg
2
40 kg
7
90 kg
3
50 kg
8
100 kg
4
60 kg
9 (MAX)
110 kg
5
70 kg
6 Maintenance
6.1 Safety Information
WARNING!
Electrical shock hazard!
– Do not remove control unit shroud.
– Refer to qualified service personnel.
– Before performing any maintenance to the
control unit, disconnect the power lead from
the wall outlet.
– Do not insert items into any openings of the
control unit. Doing so may cause fire or electric
shock by shorting the internal components.
CAUTION!
Risk of Injury or Damage
If this equipment is modified, appropriate
inspection and testing must be conducted to
ensure continued safe use of the equipment.
– Do not modify this equipment without
authorization of the manufacturer.
6.2 Cleaning and Disinfection
General safety information
CAUTION!
Risk of contamination
– Take precautions for yourself and use
appropriate protective equipment.
CAUTION!
Risk of electric shock and product damage
– Switch off the device and disconnect from
mains, if applicable.
– When cleaning electronic components consider
their protection class regarding water ingress.
– Make sure that no water splashes to the plug
or the wall outlet.
– Do not touch the power socket with wet hands.
4