Invacare® Modulite
8 Troubleshooting
8.1 Performance Troubleshooting
For additional troubleshooting information regarding the power wheelchair & electronics, refer to the Troubleshooting
section of the power wheelchair and remote user manuals (provided separately).
Symptom
Probable cause
Solutions
System tilted and/or elevated beyond
the drive lockout (DLO) angle
Return seating system to neutral (home)
position.
Drive motors not engaged
Engage drive motors.
Wheelchair power is ON, but system
does not drive
LNX powered center-mounted legrest
with telescoping footboard lowered
Retract footboard to top position.
Low batteries
Check/charge/replace batteries.
Contact your provider.
Loose/faulty electrical connection
Check cable connections/check cable ties
(too tight/too loose).
Contact your provider.
Blown fuse
Inspect/replace fuse.
Contact your provider.
Seating system not functioning
Interference/obstructions, pinched wires
Check for sources of interference or
obstructions/inspect cables for pinch
points.
Contact your provider.
70
1656515-B