WARNING
Bleached, torn, cut, frayed, or broken slings are unsafe and could result in injury. Discard Immedi-
ately.
DO NOT alter slings.
When elevated a few inches off the surface being transferred from and before moving the patient,
check again to make sure that the sling is properly connected to the attachment points of the
stand up lift. If any attachments are NOT properly in place, lower the patient back onto the surface
and correct this problem.
During transfer, with patient suspended in a sling attached to the lift, DO NOT roll stand up lift over objects
such as carpet, raised carpet bindings, door frames, or any uneven surfaces or obstacles that would
create an imbalance of the stand up lift and could cause the stand up lift to tip over. Use push handle on
the mast at ALL times to push or pull the stand up lift.
Use push handle on the mast at ALL times to push or pull the stand up lift.
Casters and axle bolts require inspection every six (6) months to check for tightness and wear.
After the first 12 months of operation, inspect all pivot points and fasteners for wear. If the metal is
worn, the parts MUST be replaced. Make this inspection every
six (6) months thereafter.
The electric motor is sealed at the factory and if service is required, the motor unit MUST BE re-
turned to the factory for repair. DO NOT attempt to open the motor or obtain local service as this
will VOID the warranty and may result in damage and a costly repair. Consult your Dealer or
Invacare for further information.
Regular maintenance of Stand Up Lift and accessories is necessary to assure proper operation.
Grounding Instructions:
DO NOT, under any circumstances, cut or remove the round grounding prong from any plug used with
or for Invacare products. Some devices are equipped with three-prong (grounding) plugs for protec-
tion against possible shock hazards. Where a two-prong wall receptacle is encountered, it is the per-
sonal responsibility and obligation of the customer to contact a qualified electrician and have the two-
prong receptacle replaced with a properly grounded three-prong wall receptacle in accordance
with the National Electrical Code. If you must use an extension cord, use ONLY a three-wire extension
cord having the same or higher electrical rating as the device being connected. In addition, Invacare
has placed RED/ORANGE WARNING TAGS on some equipment. DO NOT remove these tags. Carefully
read battery/battery charger information prior to operating, charging or servicing your stand up lift.
WEIGHT LIMITATION
DO NOT exceed maximum weight limitation of the Stand Up Lift. The maximum weight limitation is
350 lbs. regardless of any additional weight limitations on accessories.