2 Safety
2.1 General Safety Information
WARNING!
Risk of injury or damage to property
– The procedures in this manual must only
be performed by a qualified technician.
– Use only original accessories and spare
parts.
– Do not handle this product or any available
optional equipment without first completely
reading and understanding these instructions
and any additional instructional material
such as user manuals, installation manu-
als or instruction sheets supplied with this
product or optional equipment.
– After each assembly, check that all fittings
are properly tightened and that all parts
have the correct function.
WARNING!
Risk of contamination
– Clean and disinfect the product before
servicing.
IMPORTANT!
Assembly of accessories might not be
described in this service manual. Refer to
the manual, delivered with the accessory.
– Additional manuals are available on request.
See addresses at the end of this document.
IMPORTANT!
Replacement parts - for this product family
- are only available as a kit. Always use the
complete new kit when replacing a part.
– Spare parts kits can be ordered from Invacare.
Refer to your local Invacare website to access
the electronic spare parts catalogue (ESPC).
IMPORTANT!
Refer to the user manual of this product
for information on:
– Technical data
– Product components
– Labels
– Additional safety instructions
– Cleaning and disinfection instructions
The information contained in this document
is subject to change without notice.
2.2 Personal Safety Information
These safety instructions are intended to help avoid
accidents during work and must be observed under
all circumstances.
All employees coming into contact with contam-
inated products must regularly consult a company
doctor. Work clothing and personal protective equip-
ment must be available in necessary quantities and
be in proper condition. Reliable hand and surface
disinfection must be ensured.
WARNING!
Risk of contamination
– Clean and disinfect the product before
carrying out repairs.
2.3 Electrical Safety Check
IMPORTANT!
Regular electrical checks according to
EN 62353 (VDE0751) must be carried out
in line with national requirements.
Inspect all electronic components for damage to
the casing or cable, test functions and check for stray
current.