Softform Active
®
2 Rx
4.2 Replacing the Mains Cable Socket
R
efers to spare parts kit SP1633111
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
PH1 Phillips key
1. Switch off the control unit and disconnect from
mains supply.
2. Remove the bottom housing
.
3. Unplug the brown
A
and blue
B
wires
from the
defective mains socket.
Risk of malfunction or damage to the
Control Unit
Ensure the brown
A
and blue
B
wires are
plugged back into their original positions.
4. Plug the brown
A
and blue
B
wires to the new
mains socket
C
, ensuring they are installed in
the original position. Slide the socket into its slit.
5. Reinstall the bottom housing.
4.3 Replacing the User Interface
R
efers to spare parts kit SP1633233
Risk of damaging the PCB by electro-
static discharge
Handle PCB's only in Electrostatic Protected
Areas
PH1 Phillips key
1. Switch off the control unit and disconnect from
mains supply.
2. Remove the bottom housing
.
3. Remove the user's interface label
from the front
housing.
A
B
Fig. 4-8
C
Fig. 4-9
Fig. 4-10