29
PROBLEM
OTHER
SYMPTOMS
POSSIBLE
CAUSE
SOLUTION
Documentation
Motors do
not run
None
Bad connections Check all cabling. See chapter 8.6.
Fuse burnt out
Check cabling and
replace fuse
See chapters 8.6 and 8.5
Batteries
defective
Replace batteries
See chapter 8.4.2.
Cabling to power
module or
remote defective
check cabling
See chapter 8.6.
Power module
defective
Replace power
module
See chapter 8.2.2.
Corroded
contacts
Water, salt or
urine has
penetrated
Check cabling,
replace if
necessary.
See chapter 8.6.
Motor
makes
clicking
noise
None
Bearing
defective
Replacing the
motor
See chapter 8.11.1.
Collector
defective
Replacing the
motor
See chapter 8.11.1
Scraping
noise or
motor
blocked
None
Bearing
defective
Replacing the
motor
See chapter 8.11.1.
Transmission
defective
Replace
transmission.
See chapter 8.11.1.
Transmissio
n makes
clicking
noise
None
Transmission
defective
Replace
transmission.
See chapter 8.11.1.
Drive wheel
loose
Tighten drive
wheel, secure
bolts with Loctite if
necessary
See chapter 8.10.1.