SERVICE MANUAL
Invacare
®
- STORM
4
MAX
48
•
Pull the plug (1) out of the lighting PCB.
•
Undo the two Allen screws (2) with a 4 mm Allen key.
•
Lift off the lighting PCB together with the retaining bracket.
•
Loosen the two Allen screws (3) and the corresponding nuts on the rear using a 4 mm Allen
key and an 8 mm socket spanner.
•
Replace the lighting PCB.
Refitting the lighting PCB:
•
Refit the parts in reverse order.
•
Plug all the plugs into their old positions.
•
To conclude, check all the mobility aid functions.