SERVICE MANUAL
Invacare
®
- STORM
4
MAX
56
8.4.2
Replacing batteries
ATTENTION:
Injury hazard and possible material damages if batteries are handled improperly!
•
The installation of new batteries may only be carried out by authorised specialists.
•
Observe the warning information on the batteries.
•
Only use battery versions stated in the specifications.
ATTENTION:
Fire and burns hazard if battery terminal is bypassed!
•
Please take great care to ensure that the battery terminals are never short-circuited with tools
or mechanical mobility device parts!
•
Ensure that the battery terminal caps have been replaced if you are not working on the battery
terminals.
ATTENTION: Danger of crushing!
The batteries are extremely heavy. This results in injury hazards to your hands.
•
Bear in mind that the batteries are sometimes very heavy!
•
Please handle the batteries with care.
WARNING: BURN HAZARD!
Injury hazard due to discharged acid.
•
Always wear acid-proof protective gloves when handling batteries.
•
Always wear protective goggles when handling batteries.
What to do if acid is discharged:
•
Always take clothing which has been soiled by or dipped in acid off immediately!
•
Rinse any areas of your skin which has come into contact with battery acid off immediately
with plenty of water!
If contact with eyes is made:
•
Rinse the affected eye under running water for several minutes! You should also consult an
eye specialist immediately afterwards!
Requirements:
•
11 mm socket spanner
Removing the batteries:
•
Make the batteries accessible as described in chapter 8.4.1.