SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
23
8
E08
Controller Fault
00
Controller fault
Check connections and wiring. Replace
power module.
01
RAM fault
02
ROM fault
03
CPU fault
04
EEPROM fault
05
Watchdog fault
06
Stack fault
07
Software fault
08
Power-up testing fault
09
Relay fault or precharge fault
10
Bridge fault or disable all fault
11
Electronics fault: Thermistor
12
Calibration setting fault
9
E09
Communications
Fault
00
Remote connection lost
Check connections and wiring. Replace
Bus cable.
01
Low communication mode
10
E10
General Fault
00
General fault
Check all connections and wiring. Contact
Invacare Technical Service.
11
E11
Incompatible/
incorrect Remote
00
Incompatible/incorrect
Remote
Wrong type of remote connected. Ensure
the branding of the joystick matches that
of controller unit.
NUMBER OF FLASHES
DIAGNOSTICS CODE
ERROR CODE DESCRIPTION
SUB CO
DE*
DET
A
ILS
OF
ERROR CODE
POSSIBLE SOLUTION