Invacare® TDX® SP2 Series
Fig. 7-12
Fig. 7-13
WARNING!
Risk of Death or Serious Injury
Failure to observe these warnings can cause an
electrical short resulting in death, serious injury,
or damage to the electrical system.
– The POSITIVE (+) RED battery cable MUST
connect to the POSITIVE (+) battery
terminal(s)/post(s).
– The NEGATIVE (-) BLACK battery cable
MUST connect to the NEGATIVE (-) battery
terminal(s)/post(s).
– NEVER allow any of your tools and/or battery
cable(s) to contact BOTH battery post(s) at the
same time. An electrical short may occur and
serious injury or damage may occur.
– Install protective caps on positive and negative
battery terminals.
– Replace cable(s) immediately if cable(s)
insulation becomes damaged.
– DO NOT remove fuse or mounting hardware
from POSITIVE (+) red battery cable mounting
screw.
1.
Remove the batteries from the wheelchair. Refer to
7.2.12 Removing the Batteries from the Wheelchair,
page 50
.
2.
Cut the tie-wraps
A
that secure the battery terminal
covers
B
to the battery terminals.
3.
Slide the RED battery terminal cover back on the red
battery cable to expose the POSITIVE battery terminal
C
.
4.
Slide the BLACK battery terminal cover back on the Black
battery cable to expose NEGATIVE battery terminal
D
.
54
60101877-B