Technical Data
13-6
3800 Vital Signs Monitor Service Manual
Revised October 2004
SureTemp Plus Temperature
Compatibility
SureTemp Plus Series Probes
Display Units
°F and °C (user-selectable).
Measurement Resolution
0.1 °F (0.1 °C).
Measurement Range
96.0 to 105.0 °F (35.6 to 40.6 °C).
Measurement Accuracy
±0.2 °F (±0.1 °C) plus probe tolerance.
Non Invasive Blood Pressure
Method
Oscillometric.
Input Connector
Single Lumen Hose (Quick-Disconnect fitting).
Cuffs Available
Infant, Pediatric, Small Adult, Adult, Large Arm. Refer to
Section 15.
Derived Parameters
Systolic, Diastolic and Mean.
Resolution
1 mmHg.
Measurement Range
Systolic: 30 to 250 mmHg .
Diastolic: 10 to 210 mmHg.
Mean: 20 to 230 mmHg.
Measurement Accuracy
Systolic: Within 5 mmHg ±8 mmHg.
Diastolic: Within 5 mmHg ±8 mmHg.
Mean: Within 5 mmHg ±8 mmHg.
Pulse Rate Range
30 to 240 bpm.
Pulse Rate Accuracy
±5% or ±2 bpm, whichever is greater.
Update Interval
Upon measurement completion.
Measurement Time
30 seconds (typical), < 135 seconds (maximum).
Initial Cuff Pressure
160 mmHg (user-selectable over a range of 100 to 270
mmHg).
Repeated Cuff Pressure
Previous sy 40 mmHg.
Static Cuff Pressure Accuracy
±3 mmHg.
Overpressure Cutoff
290 ± 3 mmHg (normal means), 300 ± 30 mmHg (back-up).
Measurement Modes
Single Measurement or Auto (Interval) Measurement.
Auto Measurement Settings
Off, 3, 5, 10, 15, 30, 60, 120 minutes.
Recorder
Chart Speed
22 mm/second.
Width
40 mm.
Paper Type and Size
Non-Grid Thermal Paper, 50 mm wide.
Resolution
800 dots/inch.