background image

 

  SF6 Leakmate Instrument User Manual V1.0 

  

 

 

Unrivalled Gas Detect

ion.

 

Page 2 of 9 

ionscience.com

 

DECLARATION OF CONFORMITY 

 

Name und Adresse des Herstellers / Manufacturer

 ́

s name and address ISM Deutschland GmbH, Laubach 

30, D-40822 Mettmann, Germany 

 

ISM Deutschland GmbH declare the conformity of the product 

 

Bezeichnung / Product name: SF6-Lecksuchger

t / SF6-Leak Detector 

 

Typ / Type: 

 

SF6 LEAKMATE 

 

Mit den folgenden Bestimmungen / with the following applicable regulations: 

 

DIN EN 61326-1 VDE 0843-20-1

:2013-07, Elektrische Mess-, Steuer-, Regel- und Laborger

te - EMV- 

Anforderungen Gruppe 1, Klasse B (St

raussendung) / Electrical equipment for measurement, control and 

laboratory use 

 EMC requirements. Group 1, Class B equipment - (emissions section only) 

 

DIN EN 61326-1 VDE 0843-20-1

:2013-07, Elektrische Mess-, Steuer-, Regel- und Laborger

te - EMV- 

Anforderungen (St

rfestigkeit) / Electrical equipment for measurement, control and laboratory use 

 EMC 

requirements. Industrial location immunity - (immunity section only) 

 

CFR 47:2008 Class A 
Code of Federal Regulations: 15 Subpart B- Radio Frequency Devices - Unintentional Radiators. 

 

Andere Richtlinien / Other Standards 

 

EN ISO 9001: 2015 Qualit

tsmanagementsysteme

 Anforderungen / Quality Management System 

 

Requirements 

 

ISM Deutschland GmbH 

 

C. VERLEY - CEO 

 

 

 

 

 

 

 

Summary of Contents for SF6 LEAKMATE

Page 1: ...d Gas Detection ionscience com Distributed by Air Met Scientific Pty Ltd Work with Confidence Air Met Sales Service P 1800 000 744 F 1800 000 774 E sales airmet com au Air Met Rental P 1300 137 067 E...

Page 2: ...und Laborgera te EMV Anforderungen Gruppe 1 Klasse B Sto raussendung Electrical equipment for measurement control and laboratory use EMC requirements Group 1 Class B equipment emissions section only D...

Page 3: ...g 4 3 1 Battery Installation 4 3 2 Installing Replacing the Sensor Tip 4 Operation 5 4 1 Turn the detector on off 5 4 2 Recommended Leak Detection Procedures 5 4 2 1 Principle of detection 5 4 2 2 Pre...

Page 4: ...h rooms before entering Do not operate the instrument with no sensor tip in place there is risk of electric shock Commissioning 3 1 Battery Installation Hold the instrument with the probe facing downw...

Page 5: ...mination is effectively suppressed The RESET button is used to trigger zeroing manually should conditions change during detection 4 2 2 Precautions Be sure to ventilate any closed space before enterin...

Page 6: ...Sensor tip The sensor tip may get contaminated over time when a decay in sensitivity becomes apparent soak it in absolute alcohol for a few minutes then dry it carefully with compressed air Again nev...

Page 7: ...ed Clean as pointed out above The sensor tip is subject to aging and may be worn Use a new sensor tip The detector gives false alarms The ambient atmosphere s content of detectable substances may have...

Page 8: ...oltage Ionization Sensitivity approx 1E 6 mbar l s for SF6 Response Recovery Time approx 1s Audio Frequency keyed alarm sound Display 7 pos LED Bar tricolour Probe 355 mm long flexible Self Diagnostic...

Page 9: ...x 49 2104 144825 Web www ism d de Email info ism d de Italian Office ION Science Italia Via delle Querce 1 g 40011 Anzola dell Emilia Bologna ITALY Tel 39 051 0561850 Fax 39 051 0561851 Web www ionsci...

Reviews: