  SF6 Leakmate Instrument User Manual V1.0 

  

 

 

Unrivalled Gas Detect

ion.

 

Page 6 of 9 

ionscience.com

 

 Maintenance  

5.1. Cleaning  

Maintenance restricts to cleaning the instrument with a damp cloth, use mild detergent if 

required. Do not use any solvents as they may be aggressive to the case material.  

5.2. Sensor tip  

The sensor tip may get contaminated over time, when a decay in sensitivity becomes apparent 

soak it in absolute alcohol for a few minutes, then dry it carefully with compressed air. Again, 
never use solvents; they may destroy the sensor tip. Always turn the instrument off before 
removing the sensor tip!  

5.3. Batteries  

Always unload the batteries from the instrument when it is stored for extended periods of time. 

Exhausted batteries may leak aggressive acid and damage the instrument.  

  

  

 

Summary of Contents for SF6 LEAKMATE

Page 1: ...d Gas Detection ionscience com Distributed by Air Met Scientific Pty Ltd Work with Confidence Air Met Sales Service P 1800 000 744 F 1800 000 774 E sales airmet com au Air Met Rental P 1300 137 067 E...

Page 2: ...und Laborgera te EMV Anforderungen Gruppe 1 Klasse B Sto raussendung Electrical equipment for measurement control and laboratory use EMC requirements Group 1 Class B equipment emissions section only D...

Page 3: ...g 4 3 1 Battery Installation 4 3 2 Installing Replacing the Sensor Tip 4 Operation 5 4 1 Turn the detector on off 5 4 2 Recommended Leak Detection Procedures 5 4 2 1 Principle of detection 5 4 2 2 Pre...

Page 4: ...h rooms before entering Do not operate the instrument with no sensor tip in place there is risk of electric shock Commissioning 3 1 Battery Installation Hold the instrument with the probe facing downw...

Page 5: ...mination is effectively suppressed The RESET button is used to trigger zeroing manually should conditions change during detection 4 2 2 Precautions Be sure to ventilate any closed space before enterin...

Page 6: ...Sensor tip The sensor tip may get contaminated over time when a decay in sensitivity becomes apparent soak it in absolute alcohol for a few minutes then dry it carefully with compressed air Again nev...

Page 7: ...ed Clean as pointed out above The sensor tip is subject to aging and may be worn Use a new sensor tip The detector gives false alarms The ambient atmosphere s content of detectable substances may have...

Page 8: ...oltage Ionization Sensitivity approx 1E 6 mbar l s for SF6 Response Recovery Time approx 1s Audio Frequency keyed alarm sound Display 7 pos LED Bar tricolour Probe 355 mm long flexible Self Diagnostic...

Page 9: ...x 49 2104 144825 Web www ism d de Email info ism d de Italian Office ION Science Italia Via delle Querce 1 g 40011 Anzola dell Emilia Bologna ITALY Tel 39 051 0561850 Fax 39 051 0561851 Web www ionsci...