Maintenance Procedures
41
UV Filter Maintenance
(PPB Machines only)
Ultra-violet light kills biological components in the feed water such as Legionella. It also
creates low-level ozone gas. This gas has strong oxidisation properties capable of destroying
a wide range of pathogens including prions. It prevents bio-film thereby prolonging filter life.
IN ORDER TO PRESERVE THE BENEFITS OF UV/03 STERILISATION THE UV LAMP
MUST BE REPLACED EVERY 12 MONTHS
UV LAMP REPLACEMENT
DISCONNECT MAINS ELECTRICITY BEFORE PERFORMING ANY
MAINTENANCE ON THE UV UNIT
DISCONNECT THE MAINS WATER SUPPLY BEFORE PERFORMING
ANY MAINTENANCE ON THE UV UNIT
NEVER LOOK DIRECTLY AT THE ILLUMINATED UV LAMP
PROPER FUNCTIONING OF THE UV LAMP CAN OBSERVED BY A
BLUE GLOW SHOWING THROUGH A SMALL SECTION OF CLEAR
TUBING AT THE WATER CONNECTION