Maintenance Procedures – Winterising and Storage
45
DI Filter
If the DI filter is still in serviceable condition it may be left fitted to the machine. As a
precaution, remove the snap ring but leave the filter cap in place. DO NOT EXPOSE TO
FREEZING CONDITIONS. Replace the snap ring prior to putting the machine back into
service.
NGDI Filter
If the NGDI filter is still in serviceable condition it may be left fitted to the machine. As a
precaution, remove the snap ring but leave the filter cap in place. DO NOT EXPOSE TO
FREEZING CONDITIONS. Replace the snap ring prior to putting the machine back into
service.
UV/O3
To protect against any risk of frost damage, remove the UV/O3 lamp by disconnecting the
electrical connections at both ends. Withdraw the lamp followed by the quartz tube. Otherwise
DO NOT EXPOSE TO FREEZING CONDITIONS. The UV/O3 lamp should be replaced at
12-month intervals. Putting the machine back into service at the end of storage creates a
good opportunity to replace the lamp.
Battery
Ensure that the 12v battery is charged periodically during the period of storage