Troubleshooting Procedures
Condition
Solution
Low supply water flow or pressure
1. Eliminate all extension hoses and
connect directly to spigot with a
single hose.
2. Check for blockage or kinks in
hose.
3. Connect to different water source.
4. Clean or replace sediment filter.
High wastewater flow
with low
pure water flow good system
pressure
1. Remove DI cartridge and retest
2. Extremely cold water supply
3. Plugged RO membrane – Replace
RO membrane
High pure water TDS and/or high
pure water flow
1. Damaged membrane –replace
2. Membrane installed upside down
Low System pressure
with low
waste water flow
1. Verify supply water conditions
2. Clean or replace carbon sediment
filter.
3. Damaged pump – replace pump
p. 10