Interruption of Operation
Page 39
Operating manual LC 340.01-A.3.5/6
Current revision status: Apr 4, 2019
Figure 12: LaserNet
– Alarm messages page Chiller
Figure 13: LaserNet
– Alarm messages page Events
8.2.1 WATER LEVEL ALARM
Alarm indication:
The alarm lamp
“Alarms / Water level”
is lighting red. In the page
“Events”
you will
find detailed information about active alarm message:
a) Fill up tap water tank
b) Fill up DI water tank
Action:
This alarm will stop the chiller operation. Check the water level of tap and DI water and
refill.
Reset of the alarm:
After refilling tap and DI water, you can reset warning message. The red light will be
switched off.
ADVICE:
Make sure, that the temperature of the water does not exceed 30°C.
Warnings
Alarms