IPU 40420 Issue 2
Page 14 of 26
5.1.
Infrared settings
IR 32 DS
♦
Emissivity
Set the emissivity value between 0.10 and 1.00 using the left/right navigation
buttons.
Pressing Hotkey 3 (
ε
Table) gives a table of emissivity values of common materials
from which a selection can be made.
Note that if more than one cursor is selected,
individual emissivity’s can be assigned to
each cursor.
♦
Palette
Display the image using different colour palettes.
1.
Ironbow
5.
High Contrast
2.
Rainbow
6.
Rainbow 16
3.
Isotherm Style
7.
Black Hot
4.
Hot Metal
8.
White Hot
♦
Reflected Temp.
Set this to the ambient temperature to correct for reflected radiation when an
emissivity of less than 1 is selected.
♦
Temp. units
Choose between
°
C and
°
F.
♦
Integration
Chose an integration period from 1 (fast) to 4 (slow).
A longer integration time will reduce the noise in the image, but will slow down the
response to moving images.
♦
Interpolation
Choose Off or On.
This shows or hides the thermal image pixelation.