Revision 5 (p/n: 17120932)
INDUSTRIAL SCIENTIFIC
27
Set Confidence Indicator
The Set Confidence Indicator screen is used to enable the confidence indicator
during monitoring in normal instrument operation. A value of ‘0’ indicates that the
confidence indicator ‘flash’ is to be disabled. A value of ‘1’ indicates that the
confidence indicator ‘flash’ will be enabled.
On entering the screen, the primary
character display will show the present
confidence value. The rightmost
character of the primary display will
blink to indicate that it is ready to
change. The Enter button will toggle the
rightmost character between 0 and 1.
The Mode button will cause a transition
to the Set Security Code screen.
Set Confidence Indication
Actions Available from the Set Confidence Indication Screen
Action Response
Mode button press
Go to Set Security Code
Enter button press
Toggle Confidence Indication
setting
Mode and Enter buttons
held (3 sec.)
+
Go to Gas Monitoring screen
30 second timeout
Go to Gas Monitoring screen
NOTE:
When the confidence indicator is enabled, the monitor will flash every
90 seconds.
NOTE:
The factory default condition for the confidence indicator is “OFF.”
NOTE:
The backlight of the GasBadge Plus flashes as part of all alarm sequences,
except for battery low and the confidence indicator.