1. General information 

 
The crate ECH 238_1200W is able to carry up to 8 Multi Channel HV-modules of the series EHS, EDS or EBS . 
The crate provides the necessary supply voltages and connections for remote control via CAN-interface.  
 
Option 

-UPS:

 

- Integrated UPS (bridge time at least 1 min) 

 

Sufficient air flow must be ensured during the operation of the unit.  

 
 

2. Technical data 

 

 

ECH 238 M 

AC supply voltage

 

110 .. 240 V / max.12 A ( fuses double sided) 

DC supply voltages 

+ 24 V ( up to 50 A) 

+   5 V ( up to 5 A) 

Total Power 

max. 1200 W 

Floating 

 

 

max. difference of voltage between PE and internal  GND: 

 

30 V

 

clamped via  antiparallel suppressor-diode with V

Z

 = 56 V 

Mechanical layout 

 

19“ – Standard BIN 6 U / ca. 450 mm depth 

 

Module slot depth 220 mm 

 CAN-connectors: 

RJ45 

Weight 

 10 

kg 

Operating Temperature 

 

0°C … 40°C 

Storage Temperature 

 

-20°C … 50°C 

Air cooling 

 

 

In case of desk operation, rack mounting or the use of several crates in a stack 
forced air cooling must be provided. 

Please use our 1U fan-unit for rack mounting, supplemented with 
mountable feet for desk operation. 

 

Summary of Contents for ECH 238 1200W

Page 1: ...eration Attention It is not allowed to use the unit if the covers have been removed We decline all responsibility for damages and injuries caused by an improper use of the module It is highly recommended to read the manual before any kind of operation Note The information in this manual is subject to change without notice We take no responsibility for any error in the document We reserve the right...

Page 2: ...red during the operation of the unit 2 Technical data ECH 238 M AC supply voltage 110 240 V max 12 A fuses double sided DC supply voltages 24 V up to 50 A 5 V up to 5 A Total Power max 1200 W Floating max difference of voltage between PE and internal GND V 30 V clamped via antiparallel suppressor diode with VZ 56 V Mechanical layout 19 Standard BIN 6 U ca 450 mm depth Module slot depth 220 mm CAN ...

Page 3: ...ront p a Mod_addr module address b22 hard wired 31 b Bank_addr module address b23 bank switch on front p c GND a Mod_addr module address b20 hard wired 32 b Mod_addr module address b21 hard wired c GND 3 2 External CAN Bus The external CAN Bus to control the iseg HV module CAN HV and the crate with the built in CAN crate controller CAN CC is connected through RJ45 connectors on the front panel The...

Page 4: ... built in multichannel module with option _SL is active then an output voltage in any channel of it is only present if this safety loop is closed STATUS LED green red green indicates POWER ON 24V DC O K and no CAN error and Option UPS Battery on float charge STDBY AC supply ON remote control possible POWER ON ON OFF switch 24V DC Supply AC Power ECH x38 SL1 SL2 0 CAN HV POWER ON 1 BANK SELECT 2 ST...

Reviews: