S p e z i a l e l e k t r o n i k G m b H
iseg
c/o ROTECH
Email: sales@iseg-hv.de
Phone ++ 49 351 / 26 95 - 260
Spezialelektronik
Bautzner Landstr. 45
http://www.iseg-hv.com
Fax
++ 49 351 / 26 95 - 261
12
GmbH
D - 01454 Rossendorf
Germany
Group access: Voltage supplies and module temperature
(Read-write),
extended access list
Read-write
Byte
Identifier
DATA_ID
Controller (DLC = 1):
Bit
ID1
ID0
7
6
5
4
3
2
1
0
Read voltage supplies and the
Designation
EXT_
INSTR
DATA
_DIR
module temperature
Data
1
1
1
1
0
0
0
0
0
0
Description
read
⇓
Response module (DLC = 8)
Byte
Identifier
DATA_ID
DATA_n
Bit
ID1
ID0
7
6
5
4
3
2
1
0
6
5
4
3
2
1
0
Designation
EXT_
INSTR
DATA
_DIR
U1
U2
U3
U4
U5
t2
t1
Data
1
0
1
1
0
0
0
0
0
0
x
x
x
0
0
x
x
Description
write
+24V
+15V
+5V
0
0
Temperature
U1 to U3: Voltage resolution 100 mV
U4; U5 not available on EHQ 8005-F
t2 & t1: Module temperature resolution 0,1 °C
Out of range (see
Group access: General status module
) will be generated if tolerance of voltage supplies is
more than
±
5%.
Group access : Existing hardware channels
(Read-write/Write access),
extended access list
Read-write
Byte
Identifier
DATA_ID
Controller (DLC = 1):
Bit
ID1
ID0
7
6
5
4
3
2
1
0
Read the existing hardware channels
Designation
EXT_
INSTR
DATA
_DIR
at the corresponding module
Data
1
1
1
1
0
0
1
0
0
0
Description
read
⇓
Response module (DLC = 3)
Byte
Identifier
DATA_ID
DATA_1
DATA_0
Bit
ID1
ID0
7
6
5
4
3
2
1
0
7
...
0
7
...
0
Designation
EXT_
INSTR
DATA
_DIR
N3 N2 N1 N0
LSB
Data
1
0
1
1
0
0
1
0
0
0
x
Description
write
x=1: channel is existing
x=0: channel is not existing
LSB = channel 1