3.6.4 AIO connection
See section 3.5 Analogue I/O interface (AIO).
3.7 Troubleshooting
INFORMATION
INFORMATION
If these instructions do not lead to a good result, this unit must be checked by an authorised agent or shipped
to the factory.
Trouble
Solution
•
Unit does not provide output voltage.
➜
Check supply voltage
Check environmental temperature and cooling conditions
Check control
Check INHIBIT function
•
External fuses trip during switch on.
➜
Use fuses with slow characteristic (inrush current 25 A)
3.8 Maintenance
Repair and maintenance may only be performed by trained and authorized personnel.
INFORMATION
INFORMATION
For compliance of the specified accuracy of set and monitor signals, the unit has to be recalibrated once a year.
EPS High Voltage Power Supply series | Last changed on: 2020-12-02 |
www.iseg-hv.com