S p e z i a l e l e k t r o n i k G m b H
iseg Spezialelektronik GmbH
Email: sales@iseg-hv.de
Tel ++ 49 (0)351 / 26 996 - 0
Bautzner Landstr. 23
http://www.iseg-hv.de
Fax ++ 49 (0)351 / 26 996 - 21
22
D - 01454 Radeberg / Rossendorf
Module EventStatus (read/write access)
:READ:MODule:EVent:STATus?
The Module EventStatus register describes the
captured
status. Depending on the status of the mod-
ule the bits will be set but not reset. A reset has to be done by the user by writing an 1 to this event
bit. All Events can be cleared by :CONFIGURE:EVENT:CLEAR.
Bit15
Bit14
Bit13
Bit12
Bit11
Bit10
Bit9
Bit8
Res
ETMPngd
ESPLYngd
res
res
ESFLPngd
res
res
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
res
res
res
res
ESrvc
res
res
res
ETEMPngd
EventTemperatureNotGood Event: Temperature is above 55°C
ESPLYngd
EventSupplyNotGood
Event: at least one of the supplies is not good
ESFLPngd
EventSafetyLoopNotGood
Event: Safety loop is open
ESrvs
EventService
Event: A hardware failure of the HV module has been detected. The HV is
switched off without the possibility to switch on again. Please consult the
iseg Spezialelektronik GmbH.
res
Reserved