7 Output formats for voltage and current
INFORMATION
INFORMATION
The number of displayed output digits is shown in the following table.
The actual measurement resolution depends on the device hardware (standard/high precision), the ADC
sample rate and the digital filter setting.
V
nominal
Output format for voltage values
1 V ≤
V
nom
< 10V
1.23456V
10 V ≤
V
nom
< 100 V
12.3456V
100 V ≤
V
nom
< 1 kV
123.456V
1 kV ≤
V
nom
< 10 kV
1.23456E3V
10 kV ≤
V
nom
< 100 kV
12.3456E3V
Table 11: Output formats for voltage values
I
nominal
Output format for current values
10 µA ≤
I
nom
< 100 µA
12.3456E-6A
100 µA ≤
I
nom
< 1 mA
123.456E-6A
1 mA ≤
I
nom
< 10 mA
1.23456E-3A
10 mA ≤
I
nom
< 100 mA
12.3456E-3A
100 mA ≤
I
nom
< 1 A
123.456E-3A
1 A ≤
I
nom
< 10 A
1.23456EA
10 A ≤
I
nom
< 100 A
12.3456EA
Table 12: Output formats for current values
For high precision NHR and SHR (N04C2 or N06C2), there are some differences:
•
The voltage format has one digit more precision
•
The current format is the same, but a second current measurement range with the corresponding exponent exists.
Depending on the range, the exponent is E-3 or E-6.
SCPI protocol common instruction set | Last changed on: 2020-10-07 |
www.iseg-hv.com
42/61