Event Bit
Blocking
Description when Bit is set to 1
Event Voltage Bound Lower
The measured voltage has been or is lower than the programmed lower bound
Event Voltage Bound Upper
The measured voltage has been or is higher than the programmed upper
bound
Event Voltage Ramp Down
Voltage ramp down to a lower absolute value has been started
Event Voltage Ramp Up
Voltage ramp up to a higher absolute value has been started
Event Current Ramp Down
Current ramp down has been started
Event Current Ramp Up
Current ramp up has been started
Event End of Current Ramp
A running current ramp has reached its destination
Event Voltage Limit
Yes
Voltage limit V
lim
has been or is exceeded
Event Current Limit
Yes
Current limit I
lim
has been or is exceeded
Event Current Trip
Yes
Trip is set when V
lim
or I
lim
or I
set
has been exceeded in Kill Enable or I
set
has been
exceeded in Delayed Trip
Event External Inhibit
Yes
External Inhibit has been or is active
Event Voltage Bounds
Measured voltage has been or is out of bounds
Event Current Bounds
Measured current has been or is out of bounds
Event Constant Voltage
–
Constant Voltage has been or is active
Event Constant Current
–
Constant Current has been or is active
Event Emergency Off
Yes
Channel was shut down with emergency off
Event End Of Voltage Ramp
–
A running voltage ramp has reached its destination
Event On To Off
The channel was switched off due to a blocking event (Voltage/Current Limit,
Current Trip, External Inhibit or Emergency Off)
Event Input Error
–
An Input Error has occurred
SCPI protocol common instruction set | Last changed on: 2020-10-07 |
www.iseg-hv.com